Illinois EPA

Illinois environmental lab accreditation, IL ELAP and NELAP.

Which Illinois labs must be accredited, which standard applies, how out-of-state labs qualify, and the calibration findings Illinois EPA keeps writing up.

Request a demoWhat auditors check

ProgramIL ELAP
Standard2016 TNI
Rule35 Ill. Adm. Code 186
AccreditationOne year, renewable

At a glance

How environmental lab accreditation works in Illinois.

Illinois EPA’s Division of Laboratories runs the Environmental Laboratory Accreditation Program. It has been a NELAP accrediting authority since July 1, 1999, and it accredits chemistry only.

NELAP accreditation body

Program
Environmental Laboratory Accreditation Program (IL ELAP), Quality Assurance Section, Division of Laboratories, Illinois EPA
Legal basis
35 Illinois Administrative Code Part 186 (accreditation) and Part 185 (fees), under 415 ILCS 5/4
Standard
TNI EL-V1-2016-Rev 2.1 and EL-V2-2016-Rev 2.0, incorporated in 35 Ill. Adm. Code 186.115. The older Part 186 sections on calibration and QA have been repealed in favor of the TNI standard.
Who needs it
Labs submitting community water supply compliance data to the Illinois EPA Bureau of Water. Some Bureau of Land programs also require Part 186 accreditation; for example, LUST samples collected on or after January 1, 2003 must be analyzed by an accredited lab (35 Ill. Adm. Code 734.420). Other accreditation is voluntary.
Scope
Drinking water, wastewater and hazardous waste, for inorganic and organic chemistry only
Microbiology
The Illinois Department of Public Health separately certifies drinking water microbiology labs under 77 Ill. Adm. Code 465, with on-site re-evaluation every two years. That program is not NELAP.
Out-of-state labs
Secondary accreditation requires evidence of primary accreditation from a NELAP accreditation body (186.125). Illinois EPA placed a moratorium on primary accreditation of out-of-state labs on August 1, 2011. Its pages do not say whether the moratorium has been lifted, so confirm with the agency.

The accreditation cycle

The Illinois accreditation cycle.

Illinois follows the TNI standard for assessment and PT intervals rather than setting its own in the rule.

  1. Apply by method

    Apply for the drinking water, wastewater or hazardous waste methods your lab reports.

  2. New methods

    Before an on-site assessment for a new method, Illinois EPA asks for the SOP, MDLs, analyst DOCs, example data and two passing PT results at least seven days apart (Lab Communication, July 2026).

  3. One-year accreditation

    Accreditation is valid for one year and may be renewed annually (186.125).

  4. Assessment and PT

    On-site assessment and ongoing PT follow the 2016 TNI standard.

Initial calibration

What auditors check in your initial calibrations.

Illinois EPA publishes regular lab communications that list the nonconformities its assessors find most often. Calibration features in them repeatedly.

01

A range that brackets what you report

The lowest standard should sit at or below your reporting limit and the highest at or above the top concentration you report. Results outside the calibrated range need to be qualified.

02

Error at the low end, not r² alone

A correlation coefficient of 0.999 can still hide a low standard that back-calculates 30 percent off. Auditors look at relative error, especially at the lowest levels.

03

Records that rebuild the curve

Raw responses, the model and weighting used, any excluded points with the reason, and the verification that followed. An auditor should be able to reconstruct every initial calibration from the record.

04

Enough points, and no reporting above the top

A common Illinois finding is too few calibration levels: “5 points for linear and 6 points for quadratic, not counting zero as a point.” And no reporting above your highest calibration level.

Source: Illinois EPA Lab Communication, April 2026

The same communication notes that a standard at least 10 percent above the highest calibration point must be run on a defined schedule, and that the TNI requirement on measurement traceability sits in Volume 1, Module 2, section 5.6.

Self-check

Are your initial calibrations ready for an Illinois assessment?

Tick what your lab could show an auditor today. Nothing you select is saved or sent anywhere.

6Checks to work through

Select each item your lab could show an auditor today.

Request a demo

A self-assessment to help you prepare, not a compliance determination.

Illinois specifics

Assessors are telling labs where the findings are.

Illinois EPA’s lab communications are unusually direct. They list the nonconformities assessors keep finding, and initial calibration is a regular entry: curves built on too few levels, quadratic fits without the sixth point, and results reported above the highest standard because a linear dynamic range study seemed to allow it.

Each of those is a modeling decision. Choosing a quadratic fit commits the lab to more standards, and choosing a range commits it to what it can report. Making that choice consistently, and recording why, is what keeps a lab off the next list.

The program is also growing. Standard Methods 24th Edition is now accreditable, harmful algal bloom toxin analyses have been added, and Illinois enacted the Safe Public Drinking Water Act (Public Act 103-1077) in March 2025, which requires a state-only MCL for 1,4-dioxane by June 30, 2030.

How WLTR helps

Calibration evidence ready for an Illinois EPA assessment.

WLTR independently reviews the calibration your instrument software produced. Your analyst still makes the final call. What changes is that the evidence is complete, consistent across analysts, and ready before the auditor asks for it.

Calibration review today

  • Each analyte reviewed one at a time in the instrument software.
  • The model and weighting are whatever the batch was processed with.
  • r² is the number most people look at, even though it does not measure error at the low end.
  • QA signs off without an independent check on the curve.
With WLTR

  • Every analyte and every candidate model side by side in the Calibration Intelligence dashboard.
  • Diagnostics run automatically and the best fit curve is identified, with the reasons recorded.
  • Relative error at every level, including the lowest standard, shown on one screen.
  • A traceable record your QA manager can review, sign and hand to an Illinois EPA assessor.

See every candidate model for your curves, with the point counts and range each one needs, before an assessor counts them.

Request a demo

Questions

Illinois accreditation FAQs

Is Illinois a NELAP state?

Yes. Illinois EPA has been a NELAP accrediting authority since July 1, 1999, and accredits labs to the 2016 TNI standard.

Which Illinois labs must be accredited?

Labs submitting community water supply compliance data to the Illinois EPA Bureau of Water. Some Bureau of Land programs also require it, such as LUST sample analysis under 35 Ill. Adm. Code 734.420. Other accreditation is voluntary.

How many calibration points does Illinois EPA expect?

Illinois EPA lab communications cite five points for a linear calibration and six for a quadratic, not counting zero as a point.

Can a lab report results above its highest calibration standard?

Not in Illinois. Illinois EPA states that even with an established linear dynamic range, no reporting above the highest calibration level is permitted.

Who certifies drinking water microbiology labs in Illinois?

The Illinois Department of Public Health, under 77 Ill. Adm. Code 465. Illinois EPA accredits chemistry only.

Sources