BioanalysisExclusions7 min read

Excluding calibration standards: the 75% rule and the record behind it.

ICH M10 lets a laboratory reject a failing calibration standard and refit the curve, within firm limits. Each exclusion is only as defensible as the record behind it, and a pattern of exclusions is a finding in its own right.

The short version
  1. ICH M10 lets a lab reject a failing calibration standard and refit the curve, as long as at least 75% of standards and six levels still pass, and it moves the run’s LLOQ or ULOQ when an end standard is lost.
  2. An exclusion holds up when a rule written before the run made it, and the original fit, the refit, the user, the reason and any reintegration all stay in the record.
  3. The same level failing run after run usually points to the model, the weighting or the method, and calls for an investigation rather than another exclusion.

What M10 allows when a standard fails.

ICH M10, final FDA guidance since November 2022, sets the calibration test for chromatographic study runs [1][2]. Back-calculated standards should fall within ±15% of nominal, or ±20% at the LLOQ, for “at least 75% of the calibration standard concentrations, which should include a minimum of six concentration levels” (Sec. 3.3.2) [1].

With more than six levels, a failing standard “should be rejected” and the curve “re-evaluated and a new regression analysis performed” (Sec. 3.3.2). Eight levels in singlicate allow two exclusions per run; six allow none. If the LLOQ is rejected, the next lowest acceptable standard becomes the run’s lower limit and “will retain its original acceptance criteria (i.e., ±15%).” A rejected top standard makes the next acceptable one the ULOQ. The revised range should still cover the low, medium and high QCs, and study samples outside it are reanalyzed (Secs. 3.3.2, 3.3.4) [1].

If only one of the replicate LLOQ or ULOQ standards fails, the range is unchanged (Sec. 3.3.2). Validation is stricter: at least 50% of replicates per level should pass, an accuracy and precision run that loses every LLOQ or ULOQ replicate is rejected, and “if the next validation run also fails, then the method should be revised” (Sec. 3.2.4) [1].

Two rules sit behind every exclusion. The model and weighting are determined during validation (Sec. 3.2.4), and reinjecting standards “simply because the calibration standards or QCs failed,” with no identified analytical cause, “is not acceptable” (Sec. 3.3.5) [1].

Where M10 puts the exclusion record.

M10’s documentation table has no row for excluded standards; the evidence is spread out. Section 8, Table 1 asks the bioanalytical report for a “table of all runs, status (accepted and failed), reason for failure, and analysis dates,” and the analytical site for run summary tables of accepted and failed runs “including calibration curve, regression, weighting function.” For validation reports, “Values outside the acceptance criteria should be clearly marked.” Section 3.3.6 adds a list of reintegrated chromatograms with reasons [1].

FDA’s compliance program for analytical BA/BE inspections lets FDA investigators request run summary tables that can include “Failed calibrators,” and tells them, “If chromatograms for QCs and calibrators were re-integrated, determine if run acceptance was affected” (Program 7348.004, Part III) [3].

M10 ruleSec.Record that shows compliance
Acceptance criteria defined in advance3.3.2The SOP version in force for the run
±15% of nominal, ±20% at the LLOQ3.3.2Back-calculated value, %RE and limit for every standard
Failed standard rejected and curve refit; 75% and six levels still pass3.3.2Original fit and refit, both retained, with the passing count
A rejected end standard moves the range, which still covers low, medium and high QCs3.3.2, 3.3.4The run’s LLOQ and ULOQ, QCs inside them, reanalysis entries
No reinjection only because standards failed3.3.5Reinjection log naming the analytical cause
Reintegrations listed with reasons3.3.6, Table 1Reason, mode, both results and the audit trail
All runs reported with status and reason8, Table 1Run table with failed runs included

Sections are ICH M10 [1]. The right-hand column is WLTR’s mapping.

The rule has to exist before the result.

M10 makes these decisions rule-driven: acceptance criteria in the protocol, plan or SOP, reanalysis reasons predefined, and integration described a priori (Secs. 3.3.2, 3.3.4, 3.3.6) [1]. Applying a criterion fixed before the run is the guideline working as written. Choosing an exclusion after seeing which combination passes is selection.

An exclusion is defensible when the rule existed before the result. The record has to show which came first.

In its 2018 CGMP data integrity guidance, FDA prohibits sampling and testing “with the goal of achieving a specific result or to overcome an unacceptable result,” which it calls testing into compliance (Q13). Invalidating a CGMP result “requires a valid, documented, scientifically sound justification,” with the original data kept (Q2), and reprocessed chromatography needs written procedures and “each result retained for review” (Q14) [4]. It covers drug manufacturing; WLTR reads its logic as applying equally to a calibration curve.

For BA/BE studies, FDA’s April 2024 draft guidance recommends “SOPs with prespecified objective criteria” for repeat analysis, reintegration and reinjection (Sec. III.C.2.c) and audit trails that capture “when, by whom, and the reasons changes were made” (Sec. III.C.5). It names data “excluded without authorization or without detection” as a risk (footnote 14) [5]. As of October 26, 2026, it remains a draft.

FDA’s four Notifications on Data Integrity (2016 to 2025) concern falsified data or substituted samples; FDA declared affected studies unacceptable or required them to be repeated [6]. A 2021 untitled letter objected to Panexcell Clinical Lab’s removal of subject data FDA identified as concerning, “because you provided no justification for the removal” [7]. In a 2019 FDA webinar, staff described a site that reported reinjected stability results under an SOP leaving reinjection “at the discretion of the bioanalytical Principal Investigator,” without following its pre-established run acceptance criteria [8].

A recurring exclusion is a finding about the method.

Each exclusion is judged within its run; the pattern across runs says more. An unbiased LLOQ standard with a 10% CV, half its limit, fails by chance in about 4.6% of runs, so three or more failures in ten runs would happen about once in 113 sets of ten (WLTR’s calculation). A level that keeps failing usually has a systematic cause: the model, the weighting or the method.

When exclusions cluster at the low end, suspect the weighting first: an unweighted fit lets the top standards set the intercept, so the lowest standards carry large errors that shift between runs. Gu and colleagues found that “the most stable curve could be obtained when the correct weighting factor was used,” and that incorrect weighting significantly affected assay performance [9]. A one-sided bias at the LLOQ alone points instead to the standard itself, or to interference or carry-over, which M10 caps at 20% of the LLOQ response (Secs. 3.2.1, 3.2.6) [1].

M10 already reads patterns: a repeat validation failure means revising the method (Sec. 3.2.4), and failing between-run QC accuracy or precision triggers an investigation (Sec. 3.3.2) [1]. For ISR failures and trends of concern, the M10 Q&A says the investigation “should be driven by a standard operating procedure (SOP)” (Q8) [10]. WLTR recommends the same discipline for calibration standards: trend exclusions and signed %RE by level, the intercept, slope and LLOQ response, the IS response (Sec. 3.3), and reintegrations of standards by analyst and instrument.

Set the trigger in the SOP before the study, for example the same level excluded in two of five consecutive runs, which a well-behaved LLOQ does by chance in about 1.9% of five-run sets (WLTR’s calculation). If the cause is the weighting, the remedy is a method change: M10 settles the weighting in validation (Sec. 3.2.4) and treats a range change, such as raising the LLOQ, as partial validation (Sec. 3.3.3) [1]. WLTR reads refitting finished runs until they pass as the reprocessing FDA’s Q14 addresses [4].

Ten runs, one weighting decision.

Illustrative. Consider an LC-MS/MS assay in human plasma with eight levels in singlicate (1, 2, 5, 20, 100, 400, 800 and 1,000 ng/mL) and QCs at 3, 400 and 800 ng/mL. Ten runs were simulated with 1% proportional scatter plus a small floor (4.9% CV at the LLOQ) and fit unweighted, the example method’s model, and with 1/x². The example SOP excludes the worst failing standard and refits.

RunLLOQ %RE, unweighted2 ng/mL %RE, unweightedUnweighted outcomeLLOQ %RE, 1/x²
1−72.1−35.2Not acceptable, third failure−0.3
2−27.7−8.1Accepted, LLOQ excluded−2.0
3+14.2+12.5Accepted−2.3
4−55.7−30.0Not acceptable, third failure+0.7
5+32.9+17.4Not acceptable, low QC outside range−0.2
6+115.8+59.2Not acceptable, third failure−0.8
7−18.1−8.6Accepted−0.2
8−50.2−30.1Not acceptable, third failure+2.0
9+65.9+36.7Not acceptable, third failure−1.5
10−81.9−36.0Not acceptable, third failure−1.8

Illustrative data, computed by the accompanying script; %RE from the full-curve fit. Third failure: 1 and 2 ng/mL excluded, then 5 ng/mL failed. Under 1/x², no standard failed (largest |%RE| 4.3%).

The unweighted LLOQ was excluded in 8 of 10 runs, and all 15 unweighted exclusions fell at 1 and 2 ng/mL. Run 2 is the rule working as written: the LLOQ missed by −27.7% and was excluded, the refit put 2 ng/mL at −11.5% against its ±15% limit, and the run was accepted with a 2 to 1,000 ng/mL range, so its samples below 2 ng/mL go to reanalysis. The LLOQ error changed sign between runs, yet the 2 ng/mL error matched its sign in all ten at about half the size (median ratio 0.52), the signature of an intercept set by the high standards.

Figure 1 Run 2, the same responses under two weightings Linear, unweighted Linear, 1/x² Outside the M10 limit
Back-calculated accuracy (%RE)−30%−20%−10%0%+10%+20%+30%125201004008001,000Nominal concentration (ng/mL)

Illustrative data. Only the unweighted LLOQ falls outside its band.

Over 10,000 simulated runs, the unweighted LLOQ failed 78.0% of the time and 72.7% of runs were not acceptable; under 1/x² the LLOQ never failed. A method this fragile should not survive validation; the example exaggerates it so the pattern is easy to see. Refitting with 1/x² diagnoses the cause. It does not rescue the runs.

The exclusion record checklist.

An auditor will ask for each of these.

  1. The rule, dated before the run. The SOP or plan version setting the limits, exclusion procedure and minimums.
  2. The original fit. Every standard included, with every back-calculated value.
  3. The failure. Nominal and back-calculated values, %RE and the limit applied.
  4. Who, when and why. The named user, time and reason, captured by the audit trail.
  5. The refit. New parameters, back-calculated values and the count of passing standards and levels.
  6. The range consequence. The run’s LLOQ and ULOQ, the QCs inside them and samples sent for reanalysis.
  7. Integration and reinjection history. Any reintegration or reinjection of standards, with reason and both results.
  8. The trend and the review. Exclusions by level across the study, any triggered investigation, and QA review of the exclusion and audit trail before approval.

Where WLTR fits.

WLTR is a verification layer for the calibration data your CDS already produces. It reads the raw instrument export directly, records a SHA-256 hash of the original file and evaluates every candidate model against the same acceptance criteria at once, including back-calculated accuracy at every level. The full evaluation is kept, so there is no curve to shop for and no failed attempt to hide [11].

Each exclusion is recorded with the user who made it and the reason given, and it travels with the record into QA approval. Method configurations are frozen once results exist, so the model cannot quietly change. The lab sets the acceptance value for each of 15 criteria per analyte, and WLTR can trend calibration performance across batches, instruments and sites, so recurring failures show. Your validated method and SOPs still decide the reported model. WLTR verifies. Your lab validates.

Calibration verification

Keep the evidence behind every exclusion.

See how WLTR records each exclusion with its user, its reason and the full evaluation behind it.

Get Compliance Gap Assessment Download the white paper
White paper · PDF
Excluding calibration standards · Letter size · October 2026
Sources
  1. ICH, M10 Bioanalytical Method Validation and Study Sample Analysis, Step 4, May 24, 2022, Secs. 3.2.1, 3.2.4, 3.2.6, 3.3, 3.3.2 to 3.3.6 and 8 (Table 1). database.ich.org/sites/default/files/M10_Guideline_Step4_2022_0524.pdf
  2. U.S. FDA, M10 Bioanalytical Method Validation and Study Sample Analysis, Guidance for Industry, November 2022. www.fda.gov/media/162903/download
  3. U.S. FDA, Compliance Program 7348.004, In Vivo Bioavailability/Bioequivalence Studies (Analytical), implementation date May 1, 2018, Part III (Study Administration and Responsibility, B; Documentation, C). www.fda.gov/media/112533/download
  4. U.S. FDA, Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry, December 2018, Q2, Q13 and Q14. www.fda.gov/media/119267/download
  5. U.S. FDA, Data Integrity for In Vivo Bioavailability and Bioequivalence Studies, Draft Guidance for Industry, April 2024, Secs. III.C.2.c and III.C.5 and footnote 14. Listed as draft on FDA’s guidance page as of October 26, 2026. www.fda.gov/media/177404/download
  6. U.S. FDA, Notifications on Data Integrity: Semler Research (April 20, 2016); Panexcell Clinical Lab and Synchron Research Services (September 16, 2021); Synapse Labs (June 18, 2024); Raptim Research (March 28, 2025). Page content current as of July 14, 2026. www.fda.gov/drugs/drug-safety-and-availability/notifications-data-integrity
  7. U.S. FDA, CDER, untitled letter to Panexcell Clinical Lab Pvt. Ltd., issued September 15, 2021, p. 8. www.fda.gov/media/151569/download
  8. S. J. Cho and J. Kadavil, FDA CDER Office of Study Integrity and Surveillance, “Bioanalytical Inspections: Overview and Case Studies,” SBIA-BMV webinar, June 17, 2019, slides 20 and 21. www.fda.gov/media/135130/download
  9. H. Gu, G. Liu, J. Wang, A.-F. Aubry, M. E. Arnold, “Selecting the Correct Weighting Factors for Linear and Quadratic Calibration Curves with Least-Squares Regression Algorithm in Bioanalytical LC-MS/MS Assays and Impacts of Using Incorrect Weighting Factors on Curve Stability, Data Quality, and Assay Performance,” Analytical Chemistry 86(18), 2014, 8959–8966. DOI 10.1021/ac5018265. doi.org/10.1021/ac5018265
  10. U.S. FDA, M10 Bioanalytical Method Validation and Study Sample Analysis: Questions and Answers, Guidance for Industry, June 2024, Q8. www.fda.gov/media/179296/download
  11. U.S. Patent 12,038,448 B2, inventor Walter “Chip” Hinchman Sr. patents.google.com/patent/US12038448B2
WLTRWLTR, Inc. · Nevada C-Corp · 2026